A thermostatic element located at one end of the TV/SC-I (inside the enclosure) regulates the steam supply to the heater to accurately maintain the desired temperature, operating like the thermostat in an oven. Heat radiating from the steam coil heater reaches the actuator causing it to quickly shut of the steam supply when the desired temperature is reached, regardless of outside ambient. The TV/SC-I comes complete with a weather-tight bulkhead fitting for the valve body; optional bulkhead fittings for 3/8 tubing connections are available. Also available is a short configuration for installations with the valve and all connections completely within the enclosure (see TV/SC-A and ITCH product fact sheets). These economical valves are available with set points from 35°F to 200°F (1.7°C to 93.3°C)
The TV/SC-I assures accurate temperature control in an instrument or analyzer enclosure. It’s a reliable, economical alternative to costly, hazardous electric heating. This compact thermostatic control valve senses enclosure temperature and automatically regulates the low of steam to maintain the temperature. In cooling applications the reverse-acting TV/SC-IR can be used to regulate the low of glycol, water, air or other cooling media.
- Exclusive Thermoloid® thermal actuator
- Direct acting (No pilot valve)
- Stainless Steel body, fittings, spring & plug
- Compact low mass - fast response
- Corrosion resistant - long service life
- Ram-type plug for reliable tight shut-off
- Compatible with standard tubing and connections
- Maintains enclosure temperature within specified range
- Self-operating, no power or signal required
- Eliminates danger of overheating and explosion hazard
- Long service life
- Unaffected by pressure variations
- Quick and easy installation
Series | TV/SC-I |
Description | 3/8" TV/SC-I-S-SS |
Size | 3/8" NPT |
Length-in | 10 |
Length-mm | 254 |
Weight lb | 0.7 |
Weight kg | 0.3 |
Cv | 0.5 |
Maximum Operating Pressure | 200 PSIG (13.8 BAR) |
Maximum Temperature Valve End | 388°F (198°C) |
Maximum Temperature Sensing End | 150°F (66°C) over set-point limit 300ºF(149ºC) |
Opening Temperature | 35°F, 40°F, 50°F, 55°F, 60°F, 65°F, 75°F, 85°F, 90°F, 95°F, 100°F, 105°F, 110°F, 120°F, 125°F, 130°F, 140°F, 150°F, 155°F, 160°F, 170°F, 180°F, 190°F, 200°F (1.7°C, 4.4°C, 10°C, 12.8°C, 15.6°C, 18.3°C, 23.9°C, 29.4°C, 32.2°C, 35°C, 37.8°C, 40.6°C, 43.3°C, 48.9°C, 51.7°C, 54.4°C, 60°C, 65.6°C, 68.3°C, 71.1°C, 76.7°C, 82.2°C, 87.8°C, 93.3°C) |
Closing Temperature | Valve will close 10°F above the open temperature |
Body Material | Stainless Steel |
Fittings | Parker A-LOK or Swagelok |
Isometric View | Download |
Front View | Download |
Top View | Download |
Side View | Download |